Util Group

주메뉴

PRODUCT

제품소개

가스 측정기

  • >
  • 제품소개
  • >
  • 가스 측정기

[Signal Instruments] 418 Dual Beam Infra-red Gas Analyzer

Signal group, model 418, gas analyzer, 가스 분석기

418PAC-1.jpg
  • 418PAC-1.jpg
  • 제품개요
  • 기술자료
  • 액세서리
  • 프로브

상담신청하기

  • Specifications

    The Signal Instruments model 418 Infrared Gas Analyzer is designed for the continuous monitoring of CO, CO2, SO2, NO, NH3, NF3, SF6, N2O, CFCs or Hydrocarbons for research, toxic gas measurement and semiconductor gas applications.

    The gas analyzer has a totally automatic microprocessor-based operation including auto-calibration and auto-range . This makes the instrument ideal as a "smart" stand-alone unit or as a "building block" for systems integration via a RS-232 communications port or remote contact closures. Automatic diagnostics and fully-configurable alarms are also integral to the gas analyzer.

    A large multifunction, backlit LCD provides a clear display of gas concentration measurement, range, sample gas flowrate and the status of critical operating parameters including DC and bias power supplies. The intuitive, menu-driven interface and soft key buttons enable the operator to quickly configure and control the monitor's function. The unit features 19" rack mounting at 4RU height and a weight of only 38 Lb. It has a dual voltage capability of 115 Vac and 230 Vac which is switchable from the rear panel.  

     

    Data sheet

    Detector type : NDIR gas filled dual chamber single gas luft detector

    Models : NH3, SF6, NF3, CO2, CO, NO, N2O, SO2, CFCs or Hydrocarbon species

    Range : Dependant on gas, A, B, C, D or E, Auto-ranging standard

               A : 100, 500, 1000ppm

               B : 1000, 5000, 10000ppm

               C : 1, 5, 10%

               D : 5, 10, 20%

               E : 20, 50, 100%

    Response time : Typically < 15s (T5 to T95)

    Gas Flow rate : 1.0 to 5.0 LPM

    Inlet Pressure : (-5.0) to 15 psig, calibration gas : 5 to 30 psig

    Accuracy and repeatability : +/-1.0% of range or 0.2ppm

    Temperature effects : Zero : <+/-0.2% per degree C, Span : <+/-0.35 per degree C

    Detector Noise : Gas and Range dependant

    Linearity : <0.5%

    Remote control : Range, calibration and standby mode via contact closure or RS-232

    Output : 0-10Vdc and isolated 4~20mA concentration and range identifier voltage

    Logic Output : Range , calibration, gas path, mode and readiness

    Alarms : Two configurable concentration alarms, logic outputs and on fault contact set

    Power : Dual voltage 115/230Vac +/-15%, switch-selectable, 400VA max

    Environment : Temperature 5-30 degree C, 95%RH, non-condensing

    Dim : 19"rack-mount, 4RU(133.5mm) height, 550mm depth, 18Kg weight

    Warranty : 2 Years